Ground Bond Test | No-load output voltage | <12V |
Output Current Setting Range | AC:3.00-32.00A 50/60Hz 0.01A/step |
Output Current Accuracy | 3.00-9.99A:±(1.5% of setting+0.20A);10.00-32.00A:±(1.5% of setting) |
Resistance Setting Range | Upper limit:0-600mΩ;Lower limit:0-600mΩ 0=OFF |
Resistance Measurement Range | 3.00A≤I≤10.00A,0-600mΩ;10.00A<I≤20.00A,0-300mΩ;20.00A<I≤30.00A,0-200mΩ |
Compensation Resistance | 0-200mΩ |
Resistance Measurement Accuracy | ±(1.5% of reading+3mΩ) |
Test Time Setting Range | 0.0,0.1-999.9S 0.0=continuous running |
Insulation Resistance Test | Rated Output Load/Current | 1VA(1000V/1mA) 1mA |
Output Voltage Ripple | Resistance Load,1000V/1mA,<5% |
Output Voltage Setting Range | DC: 100-1000V |
Output Voltage Accuracy | ±(1.5% of setting+3V) |
Insulation Resistance Setting Range | Upper limit:0,1-3000MΩ 0=OFF; Lower limit:1-3000MΩ |
Insulation Resistance Measurement Range | 0.50-99.99-999.9-3000MΩ |
Compensation Resistance | >30MΩ |
Insulation Resistance Measurement Accuracy | 100-499V:<100MΩ:±(5% of reading+0.5MΩ);100MΩ-1000MΩ:±(10% of reading+5MΩ) |
500-1000V:<100MΩ:±(2% of reading+0.2MΩ);100MΩ-1999MΩ:±(5% of reading+2MΩ);2000MΩ-3000MΩ:±(10% of reading+5MΩ); |
Delay Time/Test Time Setting Range | 0.5-999.9s |
AC Withstand Voltage Test | Hipot Type | Cold state |
Rated Output Load | 200VA(AC:5000V/40mA) |
Output Voltage Regulation | <2%(From no-load to full load) |
Output Voltage Setting Range | AC:200-5000V ;50Hz/60Hz |
Output Voltage Accuracy | ±(1.5% of setting+3V) |
Leakage Current Setting Range | Upper limit:0.000-9.999-40.00mA ;Lower limit:0.000-9.999-40.00mA 0.000=OFF |
Leakage Current Measurement Range | 0.050-9.999-40.00mA |
Compensation Current | 0.000-2.000mA |
Leakage Current Measurement Accuracy | ±(1.5% of reading+0.005mA) |
Ramp Time/Descent time | 0.0-999.9s |
Test Time Setting Range | 0.0,0.1-999.9s 0.0=continuous running |
Leakage Current Test | Leakage Type | Dynamic, Static |
Output Voltage Measurement Range | AC:5.0-300.0V |
Output Voltage Measurement Accuracy | ±(1.5% of reading+3counts) |
Leakage Current Measuring Device | |
Leakage Current Setting Range | Upper limit:0.000-9.999-20.00mA; Lower limit:0.000-9.999-20.00mA 0.000=OFF |
Leakage Current Measurement Range | 0.050-9.999-20.00mA |
Compensation Current | 0.000-2.000mA |
Leakage Current Measurement Accuracy | ±(1.5% of reading+5counts) |
Test Time Setting Range | 0.0,0.5-999.9s 0.0=continuous running |
Power Test | Output Voltage Measurement Range | AC:5.0-300.0V |
Output Voltage Measurement Accuracy | ±(0.25% of reading+0.25% of range) |
Power Setting Range | Upper limit:0-6000W,Lower limit:0-6000W 0=OFF |
Power Measurement Range | 5.0-999.9-6000W |
Power Measurement Accuracy | ±(0.25% of reading+0.25% of range) |
Current Setting Range | Upper/Lower limit:0.000-9.999-25.00A |
Current Measurement Range | 0.050-9.999-25.00A |
Current Measurement Accuracy | ±(0.25% of reading+0.25% of range) |
Power Factor Measurement Range | 0.100-1.000 |
Power Factor Measurement Accuracy | ±(0.5% of reading+0.01) |
Delay Time Range | 0.2-999.9s |
Test Time Setting Range | 0.0,0.5-999.9s 0.0=continuous running |